Medical Product

Micro-Infusion Filter

Designed to vent air from I.V. lines and prevent bacterial contamination from reaching the patient. Available in your choice of membrane pore sizes for various applications, including critical-care neonate, pediatrics, chemotherapy, insulin-infusion and antibiotic therapy.


+ print friendly version
 


Materials of Construction:

Filter Material: 1.2 micron, Cellulose Acetate Membrane
Vent Media: .02 micron, ePTFE Membrane
Housing Material: Acrylic

Biological Safety: USP Class VI Materials
Sterilization Compatibility: Gamma Radiation, EtO

Connections:

Type: Tubing
Inlet: 0.170"od, (0.110"id w/taper lead in) Female & Male Combination
Outlet: 0.170"od, (0.110"id w/taper lead in) Female & Male Combination

Design:

Overall Dimensions: 0.850" dia x 1.51" total length
Effective Filtration Area: 1.7 cm2

Performance Data:

Bubble point: 6.0 psig
Minimum flow rate: 22 mL/min
Maximum Operating Pressure: 45.0 psi
Pressure rating: 45.0 psi
Prime time: 1.2 seconds
Bacterial retention: 1.2 micron Bacterial Retention
Filtration capacity: >500 mL with normal IV solution
Hold up volume: <0.25 mL
Usability: Disposable

Packaging:

Standard Pack: 4,000 parts/carton (bulk, non-sterile)

Specifications are for pre-sterilized products and are subject to change without notice. Results in specific applications may vary. Filtertek recommends that end user confirm fitness in specific applications.