Medical Product
Micro-Infusion Filter
Designed to vent air from I.V. lines and prevent bacterial contamination from reaching the patient. Available in your choice of membrane pore sizes for various applications, including critical-care neonate, pediatrics, chemotherapy, insulin-infusion and antibiotic therapy. Ordering Information
Product No.
0056080
Mfg. No.
0056080
Description
0.22u CA, .090" ID Ports
Get a Sample
Quantity
Add to Cart
Add to Cart
+ print friendly version
Materials of Construction:
Filter Material: 0.22 micron, Cellulose Acetate MembraneVent Media: 0.02 micron, ePTFE Membrane
Housing Material: Acrylic
Biological Safety: USP Class VI Materials
Sterilization Compatibility: Gamma Radiation, EtO
Connections:
Type: TubingInlet: 0.170"od, (0.090"id w/taper lead in) Female & Male Combination
Outlet: 0.170"od, (0.090"id w/taper lead in) Female & Male Combination
Design:
Overall Dimensions: 0.850" dia x 1.51" total lengthEffective Filtration Area: 1.7 cm2
Performance Data:
Bubble point: 40.0 psigMinimum flow rate: >3 mL/min
Maximum Operating Pressure: 45.0 psi
Pressure rating: 45.0 psi
Prime time: 1.2 seconds
Bacterial retention: 0.22 micron Bacterial Retention
Filtration capacity: >500 mL with normal IV solution
Hold up volume: <0.3 mL
Usability: Disposable
Packaging:
Standard Pack: 4,000 pieces/carton (bulk, non-sterile)Specifications are for pre-sterilized products and are subject to change without notice. Results in specific applications may vary. Filtertek recommends that end user confirm fitness in specific applications.